Salmonella 신속 검출용 간이진단키트의 성능 개선 연구
Received: Aug 28, 2012; Revised: Oct 12, 2012; Accepted: Oct 12, 2012
Published Online: Nov 30, 2012
Abatract
In this study, a rapid immuno-chromatographic assay was developed to detect Salmonella Typhimurium. This assay was based on the lateral flow immunoassay. Two different anti-Salmonella polyclonal antibody (Pab) are used in this strip assay for the double antibody sandwich immunoassay method. The strip sensor was composed of sample pad, conjugation pad, membrane, and absorbent pad. For the optimal conjugation between nano-gold and anti-Salmonella Pab, the pH value of colloidal gold solution was determined to 8.4. To prevent non-specific binding between nano gold and anti-Salmonella Pab, sample pad and membrane was treated with PBS buffer, included 2% BSA and 10% sucrose. The control and detection zones were visually detected according to antigen-antibody binding. The detection limit of the developed strip assay was 105 CFU/mL within 10 min, and improved until 104 CFU/mL after 12 hr.