Parameter | Condition | ||
---|---|---|---|
| |||
Instrument | Waters e2695 separation module | ||
Detector | Waters 2996 UV/VIS detector | ||
Wavelength | 238 nm | ||
Column | X-Bridge C18 (4.6×250 mm, 3.5 µm, Waters) | ||
Column temperature | 30°C | ||
Sample temperature | 15°C | ||
Flow rate | 0.6 mL/min | ||
Run time | 40 min | ||
Injection Volume | 20 µL | ||
Mobile phase | Time (min) | A: Water (%) | B: Acetonitrile (%) |
Initial | 95 | 5 | |
20 | 70 | 30 | |
22 | 5 | 95 | |
30 | 5 | 95 | |
32 | 95 | 5 | |
40 | 95 | 5 |